1Instruction ManualANALYTICAL PlusElectronic BalancesModelsAP110, AP210, AP310, AP250D,AP110E, AP210E, AP310E andAP250EOhaus Corporation29 Hanover Roa
10Leveling the BalanceLeveling the BalanceLeveling the BalanceLeveling the BalanceLeveling the BalanceThe balance is equipped with a level indi-cator
11PRINT *TARE *1 5 V dc (5 mA max.)2 Data Out (TXD)3 Data In (RXD)4* Tare (External signal)5 Clear To Send (CTS)6 Data Terminal Ready (DTR)7 Ground8 R
12hhmmsshhmmsshhmmsshhmmsshhmmss TIME TIME TIME TIME TIME Set current time to "hh: mm: ss", hh is between 0 - 23.TIMETIMETIMETIMETIME Print
13CommandCommandCommandCommandCommandCharacterCharacterCharacterCharacterCharacterDescriptionDescriptionDescriptionDescriptionDescriptionxSxSxSxSxS Se
14Connecting PowerConnecting PowerConnecting PowerConnecting PowerConnecting Power1. Before connecting the power cord,check that the line voltage swit
15Press and ReleasePress and ReleasePress and ReleasePress and ReleasePress and Release:Turns on the balance if it is off, zeros the balance. In the
16Symbols Used for Operation of the BalanceSymbols Used for Operation of the BalanceSymbols Used for Operation of the BalanceSymbols Used for Operatio
17USER END USER RESET YES NO ON DUAL * AL 0 OFF 1 2 ON AUTO ZERO STABILITY
18 FUNCTION FUNCTION FUNCTION FUNCTION FUNCTIONTO OPERATETO OPERATETO OPERATETO OPERATETO OPERATESETUPSETUPSETUPSET
19Turning the Balance ONTurning the Balance ONTurning the Balance ONTurning the Balance ONTurning the Balance ON1. After the Self Test in the Installa
2NOTE:NOTE:NOTE:NOTE:NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITHTHE LIMITS FOR A CLASS A DIGITAL DEVICE, PURSUANT TO PART 15 OF THE
20StabilizationStabilizationStabilizationStabilizationStabilizationBefore initally using the balance, allow time for it to adjust to its new environme
21Percent WeighingPercent WeighingPercent WeighingPercent WeighingPercent WeighingPercent Weighing is enabled only enabled only enabled only enabled o
22Parts CountingParts CountingParts CountingParts CountingParts CountingParts Counting is enabled onlyenabled onlyenabled onlyenabled onlyenabled only
23Check WeighingCheck WeighingCheck WeighingCheck WeighingCheck WeighingCheck Weighing is enabled only enabled only enabled only enabled only enabled
24Animal WeighingAnimal WeighingAnimal WeighingAnimal WeighingAnimal WeighingAnimal Weighing is enabled only enabled only enabled only enabled only en
25Fill GuideFill GuideFill GuideFill GuideFill GuideFill Guide is enabled onlyenabled onlyenabled onlyenabled onlyenabled only when Fill GuideFunction
26Fill Guide (Cont.)Fill Guide (Cont.)Fill Guide (Cont.)Fill Guide (Cont.)Fill Guide (Cont.)Reference NumberReference NumberReference NumberReference
27NOT DISPLAYED5g5g5g5g5g4g4g4g4g4g2g2g2g2g2g20g20g20g20g20g40g40g40g40g40g ORDER SAMPLES TAKEN IN5. To use the High Point function again, repeat st
28AP250D 98101-35 Sr# 3.0183User MenuAL = 1, Stb = 1AZT = OnSetup MenuLFT is OffEnabled Modes:g, dwt,ct, oz,oz t, GNcustomTael = Hong KongFunction
29Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Automatic Calibration PrintoutAutomatic Cal
3TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSINTRODUCTION ...
30Statistics PrintoutStatistics PrintoutStatistics PrintoutStatistics PrintoutStatistics PrintoutWhen statistics is enabled, a printout can be made wi
31Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Printing Data (Cont.)Percent WeighingPercent WeighingPercent Wei
32 - - - - - START - - - - -6/22/95 12:09:29 PM1 17.28667 gFill Dif 7.95202 g 2 31.75109 gFill Dif
33SETUP OPTIONS(Cont.)PC ErrorCW OptionReference TypeDisplays TypeEnd, CW Opts.AW OptionAW0, AW1, AW2,AW3Fill OptionReference TypeTarget DirectionEnd,
34MENU LOCK-OUT PROTECTIONMENU LOCK-OUT PROTECTIONMENU LOCK-OUT PROTECTIONMENU LOCK-OUT PROTECTIONMENU LOCK-OUT PROTECTIONThe menus can be locked out
35Calibration Menu ProtectionCalibration Menu ProtectionCalibration Menu ProtectionCalibration Menu ProtectionCalibration Menu ProtectionNOTESNOTESNOT
36User CalibrationUser CalibrationUser CalibrationUser CalibrationUser CalibrationUser calibration is used when it is desired to calibrate the balance
37Cal TestCal TestCal TestCal TestCal TestThis feature checks the calibration against last stored calibration information. The CalError display indic
38USER MENUUSER MENUUSER MENUUSER MENUUSER MENUThe User menu is used to adapt the balance to environmental conditions. It containssubmenus which enab
39Averaging LevelAveraging LevelAveraging LevelAveraging LevelAveraging LevelAveraging level compensates for vibration or excessive air currents. Fac
4Reference Number ... 26High Point...
40Auto-ZeroAuto-ZeroAuto-ZeroAuto-ZeroAuto-ZeroAuto-Zero minimizes the effects of temperature changes and shift on the zero reading.The balance mainta
41 SETUP MENU SETUP MENU SETUP MENU SETUP MENU SETUP MENURESETYES/NOLFTON/OFFOFFOFFOFFOFFSELECTION, UNITSggggg, dwt, ct, oz, oz t,UNIT 1 (grain), t,UN
42Setup Menu ProtectionSetup Menu ProtectionSetup Menu ProtectionSetup Menu ProtectionSetup Menu ProtectionThe Setup menu may be locked out toprevent
43Type Approved/LFTType Approved/LFTType Approved/LFTType Approved/LFTType Approved/LFTLFT can be set to ON or OFF. Selecting ON automatically sets t
44Unit Selection (Cont.)Unit Selection (Cont.)Unit Selection (Cont.)Unit Selection (Cont.)Unit Selection (Cont.)TaelsTaelsTaelsTaelsTaelsIf taels are
45StatisticsStatisticsStatisticsStatisticsStatisticsStatistics provides printed display data of: Standard Deviation either population orsample, Mean,
46NetNetNetNetNetWeight shown on the display can be referred to as a zero value (gross value) or tarevalue (net value). When enabled the display valu
47Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection
48Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection (Cont.)Custom Unit or Volume Selection
49Parts Counting ErrorParts Counting ErrorParts Counting ErrorParts Counting ErrorParts Counting ErrorParts counting Error is enabled only whenthe Par
5SETUP MENU ... 41Setup Menu Protection ...
50NORMAL DISPLAYSNORMAL DISPLAYSNORMAL DISPLAYSNORMAL DISPLAYSNORMAL DISPLAYSNONE DISPLAYSNONE DISPLAYSNONE DISPLAYSNONE DISPLAYSNONE DISPLAYSSIGN DIS
51Check Weighing Options (Cont.)Check Weighing Options (Cont.)Check Weighing Options (Cont.)Check Weighing Options (Cont.)Check Weighing Options (Cont
52Animal Weighing OptionsAnimal Weighing OptionsAnimal Weighing OptionsAnimal Weighing OptionsAnimal Weighing OptionsAnimal Weighing Options is enable
53Fill Option (Cont.)Fill Option (Cont.)Fill Option (Cont.)Fill Option (Cont.)Fill Option (Cont.)5. . If REF WT is selected, a sample reference
54Time (Cont.)Time (Cont.)Time (Cont.)Time (Cont.)Time (Cont.)6. .7. first two digits are flashing.8. or to change flashing digits to current lo
55DateDateDateDateDateDate is a feature which enables the balance to be set to a U.S.A. date standard orEuropean date standard. U.S. standard has the
56LockswitchLockswitchLockswitchLockswitchLockswitchLockswitch enables you to lock out one or more menu selections. Each menu can beindividually lock
57PRINT OPTIONSAuto PrintCont, interval,on stb,offoffoffoffoffInitialize Auto PrintInterval, non print lo,non print hiStable DataON/OFFOFFOFFOFFOFFNum
58ResetResetResetResetResetThis submenu enables you to reset allallallallall Print menu selections to the factory defaultsettings shown below. Reset
59Data BitsData BitsData BitsData BitsData BitsTo set the number of data bits to 7 or 8:1. Access thesubmenu.2..3. or .4..Baud RateBaud RateBaud Ra
6Print Stable Data Only ... 63Print Numeric Data Only ...
60ParityParityParityParityParityParity can be set to Odd, Even or None. The default setting is None. To set parity,proceed as follows:1. Access the
61Good Laboratory Practice (GLP) (Cont.)Good Laboratory Practice (GLP) (Cont.)Good Laboratory Practice (GLP) (Cont.)Good Laboratory Practice (GLP) (Co
62InitializeInitializeInitializeInitializeInitializeThis submenu allows you to:• Specify a time interval (in seconds) for automatic output.• Exclude a
63Initialize (Cont.)Initialize (Cont.)Initialize (Cont.)Initialize (Cont.)Initialize (Cont.)11. After the last digit is entered, is displayed again.1
64ReferenceReferenceReferenceReferenceReferenceWhen the Reference function is set ON, prints the value of weight used as a referencein either Check We
65ListListListListListThis submenu can be used to output a listing of current menu settings via the RS232interface. When selected, all menu settings f
66CARE AND MAINTENANCECARE AND MAINTENANCECARE AND MAINTENANCECARE AND MAINTENANCECARE AND MAINTENANCETo keep the balance operating properly, the hous
67Error Codes ListError Codes ListError Codes ListError Codes ListError Codes ListThe following list describes the various error codes and which can a
68Error Codes List (Cont.)Error Codes List (Cont.)Error Codes List (Cont.)Error Codes List (Cont.)Error Codes List (Cont.)System ErrorsSystem ErrorsSy
69SERVICE INFORMATIONSERVICE INFORMATIONSERVICE INFORMATIONSERVICE INFORMATIONSERVICE INFORMATIONIf the Troubleshooting section does not resolve or de
7INTRODUCTIONINTRODUCTIONINTRODUCTIONINTRODUCTIONINTRODUCTIONThis manual covers Installation, Operation and Troubleshooting for the OhausANALYTICAL Pl
70MODELMODELMODELMODELMODELAP110AP110AP110AP110AP110AP210AP210AP210AP210AP210AP310AP310AP310AP310AP310AP250DAP250DAP250DAP250DAP250DCapacity (g) 110 2
71LIMITED WARRANTYLIMITED WARRANTYLIMITED WARRANTYLIMITED WARRANTYLIMITED WARRANTYOhaus products are warranted against defects in materials and workma
72Ohaus Corporation29 Hanover Road,Florham Park, NJ 07932, USATel: (973) 377-9000,Fax: (973) 593-0359With offices worldwide.P/N 78209-01 R1099 ©
TABLE OF CONTENTS UNPACKING...6 INSTALLATION...
Parts Counting Error Level ...33 Parts Counting Alternate Display...
8UNPACKINGUNPACKINGUNPACKINGUNPACKINGUNPACKINGYour ANALYTICAL Plus balance was shipped with the following items:• Pan assembly (including pan, shield,