i5000 Series IndicatorsInstruction ManualT51XW IndicatorT51P Indicator
EN-8 5000 Series Indicators1.2 Overview of Parts and Controls (Cont.)Item Description1 Sense Jumper W12 Alternate Load Cell Terminal Block J43 Sen
EN-95000 Series Indicators1.2 Overview of Parts and Controls (Cont.)123456789101112131415171819202122232425No. Designation1 UNDER LED2 ACCEPT LED3
EN-10 5000 Series Indicators1.3 Control FunctionsButtonPrimary Function (Short Press)ON/ZEROTurns the Indicator on.If Indicator is On,
EN-115000 Series Indicators2. INSTALLATION2.1 Unpacking2.2 External Connections2.2.1 Scale Base with Connector to T51POhaus bases with a co
EN-12 5000 Series Indicators2.2.6 Mounting BracketPosition the wall bracket over the threaded holes in the side of the indicator as shown in Figures
EN-135000 Series IndicatorsPin ConnectionJ7-1 RTSJ7-2 TXDJ7-3 RXDJ7-4 CTSJ7-5 GND2.3.3 RS232 Interface Cable to T51XWPass the optional RS232 cable t
EN-14 5000 Series Indicators3. SETTINGS3.1 Menu StructureTABLE 3-1. MENU STRUCTURE.CALIBRATIONSETUPREADOUTMODEUNITGMPPRINT1•ZERO1)•R
EN-155000 Series Indicators3.2 Menu Navigation3.3 Calibration Menu When CAL is displayed, press the Yes button to accept the Calibration menu s
EN-16 5000 Series Indicators3.3.1 Zero CalibrationZero calibration uses one calibration point. The zero calibration point is established with no w
EN-175000 Series Indicators3.3.3 Linearity Calibration Linearity calibration uses 3 calibration points. The full calibration point is established
EN-18 5000 Series Indicators3.3.6 Geographical Adjustment Factor Refer to Table 3-2 and set the GEO factor that corresponds to your location. 00
EN-195000 Series Indicators Elevation above sea level in meters 0 325 650 975 1300 1625
EN-20 5000 Series IndicatorsReset No,YesRange Single, DualFull Scale Capacity 1…999950Graduation 0.00001…1000Power On unit
EN-215000 Series Indicators•••3.4.5 Power On UnitSet the unit of measures displayed at startup AUtO = last unit in use when turned off
EN-22 5000 Series Indicators3.4.7 Auto-TareSet the Automatic Tare functionality. OFF = Automatic Tare is disabled. ON = t
EN-235000 Series Indicators3.4.10 Beeper VolumeSet the beeper volume. OFF = disabled. LOW = soft HI = loud.3.4.11 B
EN-24 5000 Series Indicators3.5.3 FilterSet the amount of signal filtering. LOW = less stability, faster stabilization time (<1 se
EN-255000 Series Indicators3.5.5 BacklightSet the display backlight functionality. OFF = always off. ON = always on. AUtO
EN-26 5000 Series Indicators3.6.2 Weighing ModeSet the status. OFF = Disabled ON = Enabled3.6.3 Parts Counting ModeSet the stat
EN-275000 Series Indicators3.6.7 Check Weighing Mode Set the status. OFF = Disabled ON = Enabled3.6.8 End ModeAdvance to the next
EN-15000 Series IndicatorsTABLE OF CONTENTS1. INTRODUCTION ...
EN-28 5000 Series Indicators3.7.5 Ounce UnitSet the status. OFF = Disabled ON = EnabledNOTE: Ounce Unit is not available when Range i
EN-295000 Series Indicators3.7.9 End UnitAdvance to the next menu. 3.8 GMP MenuEnter this menu to set the data for Good Manufacturing Practice. D
EN-30 5000 Series Indicators3.8.3 Date SetSet the date. 00 to 99 = year position 01 to 12 = month position 01 to 31 = da
EN-315000 Series Indicators3.8.6 User IDSet the user identification. 000000 to 999999Refer to Section 3.2 Menu Navigation to enter settings.3.8
EN-32 5000 Series IndicatorsReset No,YesStable Only Off, OnAuto Print Off, On Stable (-> Load, Load and Zero), Inte
EN-335000 Series Indicators3.9.4 Print Content Sub-menuThis sub-menu is used to define the content of the printed data.ResultSet the status.
EN-34 5000 Series IndicatorsProject IDSet the status. OFF = Disabled. ON = the Project ID is printed.Scale IDSet the status. OFF
EN-355000 Series IndicatorsLine FeedSet the paper feed. LINE = move paper up one line after printing 4.LINE = move paper up four
EN-36 5000 Series Indicators3.10.2 BaudSet the Baud rate. 300 = 300 bps 600 = 600 bps 1200 =1200 bps 2400 = 2400 bps 480
EN-375000 Series Indicators3.10.7 Alternate Command Sub-menuEnter this sub-menu to set a different command character for the P (Print), T (Tare) and Z
EN-2 5000 Series IndicatorsTABLE OF CONTENTS (Cont.) 3.4.11 Beeper Signal ...
EN-38 5000 Series Indicators3.11.2 External InputSet the function to be controlled by an optional external input device such as a foot switch. O
EN-395000 Series IndicatorsContactSet the timing of the relay contacts. SIM = relays open or close at the same time. b-b-M =
EN-40 5000 Series Indicators3.12.3 Lock SetupSet the status. OFF = Setup menu is not locked. ON = Setup menu is locked.3.12.4
EN-415000 Series Indicators3.12.11 Lock GMPSet the status. OFF = GMP menu is not locked. ON = GMP menu is locked.3.12.12 Lock
EN-42 5000 Series Indicators3.13.5 Lock Print ButtonSet the status. OFF = Print button is unlocked. ON = Print button is locked
EN-435000 Series Indicators4. OPERATION4.1 Turning Indicator On/Off To turn the Indicator on, press the ON/ZERO Off button. The Indicator perfo
EN-44 5000 Series Indicators4.8 Application ModesPress and hold the FUNCTION Mode button until the desired application mode appears. Only modes ena
EN-455000 Series IndicatorsEstablishing a New APWPress the No button to increment the sample size. Choices are 5, 10, 20, 50 and 100. Toestablisht
EN-46 5000 Series Indicators4.8.4 Check WeighingUse this mode to determine if the weight of a sample is within prescribed limits.Checkweighing Limi
EN-475000 Series Indicators4.8.5 Dynamic WeighingUse this mode to weigh moving or oversized objects. The weight is held on the display until reset
EN-35000 Series IndicatorsTABLE OF CONTENTS (Cont.) 3.9.5 Layout Sub-menu ...
EN-48 5000 Series Indicators5. SERIAL COMMUNICATION The T51P and T51XW Indicators include an RS232 serial communication interface. The setup of RS
EN-495000 Series Indicators5.2 Output Format5.3 PrintoutsThe following sample print outs are generated by the Print button, “P” Command or alterna
EN-50 5000 Series Indicators Dynamic Mode Printout Check Weighing Mode Printout Calibration Te
EN-515000 Series Indicators6. LEGAL FOR TRADEWhen the indicator is used in trade or a legally controlled application it must be set up, verified an
EN-52 5000 Series Indicators Figure 6-3. T51P Paper Seal Figure 6-4. T51XW Pap
EN-535000 Series Indicators7. MAINTENANCE CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.7.1 Model T51P Cleaning
EN-54 5000 Series Indicators TABLE 7-1. TROUBLESHOOTING (Cont.).SYMPTOM PROBABLE CAUSE(s) REMEDYError 9.5 Calibration data not pr
EN-555000 Series Indicators8. TECHNICAL DATA8.1 SpecificationsMaterialsT51XWHousing:stainless-steelT51PHousing:ABSplasticDisplaywin
EN-56 5000 Series Indicators8.2 Accessories and OptionsTABLE 8-2. OPTIONS.DESCRIPTION PART NUMBERAC Relay Kit 80500720Base Mount Kit , T51P 805007
EN-575000 Series Indicators8.3 Drawings and DimensionsFigure 8-1. T51P Indicator Overall Dimensions with Mounting Bracket.Figure 8-2. T51XW Indica
EN-4 5000 Series Indicators4. OPERATION ...
EN-58 5000 Series Indicators8.4 Compliance Compliance to the following standards is indicated by the corresponding marking on the product.Marking Stan
EN-595000 Series IndicatorsWeighing Instruments verified at the place of manufacture bear one of the preceding marks on the pack-ing label and the gre
LIMITED WARRANTYOhaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warran
Ohaus Corporation 7 Campus DriveSuite 310 Parsippany, NJ 07054, USATel:(973)377-9000Fax:(973)944-7177www.ohaus.comP/N 80251400 E © 2013 Oh
EN-55000 Series Indicators1. INTRODUCTIONThis manual contains installation, operation and maintenance instructions for the T51P and T51XW Indicators.
EN-6 5000 Series Indicators1.2 Overview of Parts and ControlsFigure 1-1. T51P Indicator.123Item Description1 Data Label2 Front Housing3 Control Pa
EN-75000 Series IndicatorsItem Description1 Data Label2 Front Housing3 Control Panel4 Adjusting Knob (2)5 Mounting Bracket6 Screw (4)7 Rear housing8 D